GAS BLENDERS & GAS GENERATORS
Control, Calibration and Validation of Chemical Gas Analyzers.
![]()
Permeation tubes :

A constant flow " d " of pure gas is delivered from CALIBRAGE's sources, they are of three different types : permeation, diffusion and saturation. Swept by a carrier gas flow " D " (see Calibrators), you can generate a Calibration blend at a known concentration " C " . " C " is then calculated by the ratio " d " / " D ".
Substance list
A sealed reservoir, part of which is a polymeric membrane, is filled with pure gas or liquid . By maintaining this reservoir at a constant temperature, a very stable and uniform flow " d " of pure compound will be released from the permeation tube . This is called the permeation rate.
Substances : SO2, NO2, H2S, THT, H2O, BTX, COV ..... .....and each product with a boiling point between -80°C et +120°C.
Concentration ranges :
1 ppb-v to 30 ppm-v.
A capillary tube or a very small orifice is connected to a reservoir filled with pure liquid . By maintaining temperature and pressure constant, the tube will provide a very stable flow " d " called diffusion rate.
Substances : BTX, COV, BTX, COV, Mercury, H2O, ..... and each liquid with a boiling higher than 50°C..
Concentration ranges : from 100 ppb-v to 1000 ppm-v.
A pure liquid filled container is maintained at a constant temperature and pressure . If you sweep this container with a carrier gas you'll get a constant loss of pure product " d " due to its vapor pressure.
Substances : BTX, COV, H2O, .... and each liquid with a boiling point higher than 50°C.
Concentrations ranges : 100 ppm-v au %
Metrology of sources :
- The pure gas flow " d " from the source is measured by successive weighing within known time intervals.
- Referring to national standards of temperature, weight and time enables us to provide you traceable sources.
- accuracy on flow " d " are in the ranges +/-2 and +/-10%, depending on type of source and its operating conditions.
Note : accuracy given are always given with a confidence interval of 95% (2 x Relative Standard Deviation , RSD).
High concentrated blends or pure gas in cylinder :
We are able to provide you with a wide range of cylinders through several gas supplier .

Ask for our technical documentation :